Pro-Dent CT mk II

pro-dent_ct_mk_ii_01_750x500
pro-dent_ct_mk_ii_01_750x500pro-dent_ct_mk_ii_02_750x500pro-dent_ct_mk_ii_03_750x500pro-dent_ct_mk_ii_04_750x500pro-dent_ct_mk_ii_05_750x500pro-dent_ct_mk_ii_06_750x500pro-dent_ct_mk_ii_07_750x500pro-dent_ct_mk_ii_08_750x500pro-dent_ct_mk_ii_09_750x500pro-dent_ct_mk_ii_10_750x500
  The new Pro-Dent CT mk II phantom is a versatile quality control tool of dental Cone-Beam CT, Dental Volume Tomography (DVT) and other 3D imaging devices according to the Radiation Protection Report no 172 by SEDENTEXCT. The phantom consists of a main PMMA cylinder that houses modules with diffe ...Read more



  • Description
  • Specifications

 

The new Pro-Dent CT mk II phantom is a versatile quality control tool of dental Cone-Beam CT, Dental Volume Tomography (DVT) and other 3D imaging devices according to the Radiation Protection Report no 172 by SEDENTEXCT.

The phantom consists of a main PMMA cylinder that houses modules with different test objects. Thanks to this design, you can perform tests with devices with a small FOV at different positions in the 160 mm phantom.

 

General

  • Diameter: 160 mm
  • Made of PMMA (1.19 g/cm³)
  • Total length: 170 mm
  • Positioning aids on the outside surface of the phantom

 

Noise / uniformity section

  •     Uniform PMMA part of the phantom

 

5 layer modules section with modules containing

  • Linear Spread Function (LSF) PTFE / PMMA interface
  • Point Spread Function (PSF) - 0.25 mm tungsten steel wire in air
  • XY high contrast resolution (aluminum/polymer) - 1.0, 1.7, 2.0, 2.5, 2.8, 4.0, 5.0 LP/mm
  • Z high contrast resolution (aluminum/polymer) - 1.0, 1.7, 2.0, 2.5, 2.8, 4.0, 5.0 LP/mm
  • 6 low-contrast groups of 1.0, 2.0, 3.0, 4.0, 5.0, 6.0 mm rods made of: aluminum, PTFE, POM-C (delrin), PE-300, air and water emulating epoxy; background made of PMMA
  • 10 mm rods made of aluminum, PTFE, POM-C (delrin), PE-300, air and water emulating epoxy suspended in PMMA - pixel intensity / HU values samples
  • Stacked 20 mm diameter contrast-to-noise discs made of aluminum, PTFE, POM-C (delrin), PE-300 and air suspended in PMMA
  • Three 5.0 mm diameter titanium rods embedded in PMMA - beam hardening artefacts

 

Slice geometry section

  •     20 mm thick module with two pairs of aluminum ramps and empty rods in vertexes of a square for detailed evaluation of slice thickness and geometry and laser position verification

 

Geometric distortion section

  •     An array of 2.0 mm diameter, 3.0 mm long holes uniformly pitched at 10.0 mm intervals

 

Accessories

  • Test stand with spirit level for accurate placing of the phantom in the test position
  • Folding base for test stand positioning on the X-ray unit’s chair
  • Convenient, portable case for storing and transporting the phantom
  • Optional additional homogeneity disc

 

 

Product features:

  • Complies with:
    • Radiation Protection no 172 report by the SEDENTEXCT
    • IEC 61223-3-4 and IEC 61223-3-5
  • CE certified
  • The manual provides detailed guidelines for carrying out each test, results assessment and registration