Contact
Industrial
Physics
Radiotherapy
Radiology
Ultrasound
Home

Invenia UBUS Phantom

551m-750x500
  Designed for Use with GE’s Invenia ABUS Scanner The Model ATS UC-551M, a modified version of the ATS Model 551 Small Parts Phantom, is designed for use with GE’s Invenia ABUS scanner. The scan surface has a 17.5-inch radius of curvature, which matches the shape of the Invenia ABUS scan head. Th ...Read more



  • Description
  • Specifications

 

Designed for Use with GE’s Invenia ABUS Scanner

The Model ATS UC-551M, a modified version of the ATS Model 551 Small Parts Phantom, is designed for use with GE’s Invenia ABUS scanner. The scan surface has a 17.5-inch radius of curvature, which matches the shape of the Invenia ABUS scan head.

The rubber-based tissue mimicking material has a sound velocity of 1450 m/sec and an attenuation of 0.5dB/cm/MHz at room temperature (23°C).To ensure measurement accuracy, CIRS incorporates a thermometer strip affixed to the outside surface of the phantom housing.

The phantom is packaged in a carrying case providing added protection during transporting and storage. GE-recommended QA test procedures for the Invenia ABUS and Invenia ABUS 2.0 scanners are provided in separate documents.

 

Performance measures

  • Axial Measurement Accuracy
  • Lateral Measurement Accuracy
  • Axial and Lateral Resolution
  • Contrast Resolution
  • SNR: Signal to Noise Ratio

 

 

Dimensions 28.5 cm x 12 x 10.6 cm
Weight 3.6 kg
Housing Material PVC
Scan Surface Dimensions 26.7 cm x 10 cm curved radius of 44.5 cm
Tissue-Mimicking Material Urethane Rubber
Urethane Properties Freezing Point: -40° C
Melting Point: Above 100° C
Speed of Sound: 1450 m/s at 23°
Attenuation: 0.5 dB/cm/MHz (measured at 3.5 MHz)
Line Targets Material: Monofilament nylon
   
Model ATS UC-551M Includes
 
1x Invenia ABUS Phantom
1x Carry Case
1x User Guide
 1x GE QA Test Procedures
1x Certificate of Compliance

 

Vertical Distance Group (12 targets)

  • Depth range 0.5 - 5 cm
  • Spacing 0.5 cm

Horizontal Distance Group (2 Groups)

  • Depths Linear: 2 & 4cm
  • Sector: 2.5 & 4.5 cm
  • Spacing 0.5 cm

Axial and Lateral Resolution Groups (2 targets)

  • Target sets per depth 10
  • Spacing 0.25, 0.5, 1, 2, 3 mm
  • Depth 2, 4, 5 cm

Anechoic Stepped Cylinders (4 groups)

  • Type Non-echogenic, cylindrical
  • Number of targets 6
  • Diameter 1, 2, 3, 4 mm
  • Depth 1 - 6 cm
  • Spacing 1 cm center to center

Grayscale Targets (6 targets)

  • Type Echogenic, cylindrical
  • Diameters 6 mm
  • Spacing 1 cm center to center
  • Depth 4 cm
  • Contrast (dB) -15, -6, -3, +3, +6, +15