Contact
Industrial
Physics
Radiotherapy
Radiology
Ultrasound
Home

Peripheral Vascular Doppler Flow Phantom

ats-524-750x500
  The ATS Model 524 tissue mimicking Doppler flow phantom contains four flow channels simulating superficial vasculature. The simulated vessels are located 15 mm below the scan surface. Built-in scanning wells are provided to permit the use of water or a low viscosity gel as acoustic coupling agent ...Read more



  • Description
  • Specifications

 

The ATS Model 524 tissue mimicking Doppler flow phantom contains four flow channels simulating superficial vasculature. The simulated vessels are located 15 mm below the scan surface. Built-in scanning wells are provided to permit the use of water or a low viscosity gel as acoustic coupling agents. The phantom has four constant diameter channels, 2, 4, 6 and 8 mm.

When used with Model 769, Doppler Flow Pump, the following tests can be performed:

  • Sensitivity
  • Flow Velocity
  • Maximum Penetration
  • Flow Location

 

 

Dimensions 22 x 14 x 10 cm
Weoght 6.1 kg
Housing Materials PVC
Scan Surfaces 1
Maximum Fluid Pressure 15 psi (1.05 Kg/cm)
Scan Surface Dimensions 17.5 x 9.8 cm
Connections Quick Disconnect
Tissue-Mimicking Material Urethane Rubber
   
Model ATS 524 Includes  
1x Peripheral Vascular Doppler Flow Phantom
1x User Guide
1x Certificate of Compliance
   
Urethane Properties Freezing point: -40°C
Melting point: Above 100° C 
Speed of Sound: 1450 m/s ±1.0% at 23° 
Attenuation Coefficient:  0.5 dB/cm/MHz ± 10%
Flow Channels Type: Circular 
Number of Channels: 4 