The Next Generation Electrometer

Built on the MAX 4000's legacy of highly accurate, reference grade measurements, combined with a revolutionary touchscreen interface and a host of advanced measurement features, the SuperMAX redefines what an electrometer should be.

 

Two Independent Measurement Channels

  • Independent control over range, bias voltage, and applied system factor.
  • Both channels feature extensive range (0.001 pA to 500.0 nA, 0.001 pC to 999.9 µC) for use in a variety of applications, including ratio measurements.

 

Built-In Detector Library

  • Eliminates the need for a PC or to apply system factor and temperature/pressure corrections.
  • Store over 100 detectors and calibrations.
  • Apply individual factors to one or both measurement channels.

 

Easy-to-Use Interface

  • Large, color display with built-in touchscreen.
  • On-screen keypad and pull down menus allows for easy access to settings and detector library.

 

Store Data Collection

  • Measurements are stored from all data collection. 
  • Set automatic repeat measurements.

 

Exradin W1 Scintillator Integration 

  • Two dedicated modes in the SuperMAX allow for seamless calculations using the Exradin W1 Scintillator.

 

Specifications                                              
 
REF Number 90018
Height 8.1 cm (3.2 in)
Width 26.7 cm (10.5 in)
Length 21.1 cm (8.3 in)
Weight                                  2.4 kg (5.3 lbs)
Display Range RATE:
Low Range 0.001 pA – 500.0 pA, 1 fA resolution
High Range 0.001 nA – 500.0 nA, 1 pA resolution

CHARGE:
Low Range 0.001 pC – 999.9 μC, 1 fC resolution
High Range 0.001 nC – 999.9 μC, 1 pC resolution
Charge Collections TRIGGER:
Automatic start, stop, reset and save data based on user defined thresholds (Start: 0.2 – 9.9 pA; Stop: 0.1 – 9.8 pA)

TIMED:
User set duration (Range: 0 – 600 seconds;
Increment: 1 second)

CONTINUOUS:
Unlimited duration with manual stop
Real Time Clock Date and time stamp for all measurements for easy identification
Internal Memory Store preferences, >100 sources, >100 chamber/system factors
Range Switching User selectable — High or Low
Repeatability ± 0.1% (IEC 60731 requirement: ± 0.5%)
Long-Term Stability 0.5% (over one year)
Stabilization Time ± 0.1% (IEC 60731 requirement: ± 0.5% of value at 1 hr for measurements taken at 15 min and 6 hrs)
Response Time < 3 s on high range (IEC 60731 requirement: < 3 s)
Non-Linearity ± 0.25% (IEC 60731 requirement: ± 1.0%)
Conformity  CE 93/42/EEC
Reference class according to IEC 60731
Zero Drift ± 0.25% (IEC 60731 requirement: ± 0.5%)
Zero Shift ± 0.25% (IEC 60731 requirement: ± 0.5%)
Display 6.4" color TFT, touchscreen
Input (2) BNC two lug, triaxial connector
Bias Voltage Nominal ± 1000 volt bias

USER SETTINGS:
– 1000 to – 100, 0; 100 to 1000 (set in 1 volt increments)

ACCURACY:
± 0.3 volt
Power 100-240 VAC, 0.5 A max, 50/60 Hz input to external power supply, 9 VDC, 1.7 A power supply output to electrometer input, UL/TUL listed power supply
Zeroing Automatic zero function, user activated
Output (2) USB ports, Isolated RS-232; bidirectional 19,200 baud rate; 8 data bits; no parity; 1 stop bit;
Options SuperMAX Accessory Kit (REF 72245)
Includes extra stylus, extra USB flash drive, and set of 5 extra screen protectors