DoseView 1D

doseview-1d_600x600



Deprecated: implode(): Passing glue string after array is deprecated. Swap the parameters in /home/clients/844ee2187cee70a16a19090345f717c4/web/medical-physics/plugins/content/jw_allvideos/jw_allvideos.php on line 73

Easily acquire depth-dose measurements in an intuitive, TG-51 and TRS-398 compliant 1D water scanning system

 

0.05 mm Accuracy for Consistent QA

  • Superior 0.05 mm accuracy of position and repeatability over entire travel distance produced by leadscrew shaft.

 

Easy Operation with Handheld Pendant

  • Establish origin position and return to origin with the press of a button.
  • Three convenient positioning modes: fast, slow and step (customizable from 0.01 mm to 100 mm).
  • One-touch toggle between cm/mm.

 

Water-Equivalent Bracket Fits Range of Detectors

  • Adaptable chamber bracket fits most thimble ion chambers and rigid stem parallel plate chambers.

 

Phantom Available in Two Sizes:

  • 32 cm x 28 cm x 36 cm
  • 40 cm x 38 cm x 36 cm

 

Specifications
Dimensions  
1D Scanning Arm Height: 48.26 cm (19.0 in)
Width: 6.77 cm (2.67 in)
Length: 9.44 cm (3.72 in)
Water Tank (REF 91800) Height: 36 cm (14.17 in)
Width: 28 cm (11.81 in)
Length: 32 cm (12.60 in)
Water Tank (REF 91810) Height: 36 cm (14.17 in)
Width: 38 cm (15.75 in)
Length: 42 cm (16.54 in)
Water Tank Material Clear acrylic 0.95 cm (0.375 in)
Weight 1D Scanning Arm 1.36 kg (3 lbs)
Handheld Controller 0.23 kg (0.5 lbs)
Water Tank (ref 91800, empty) 6.35 kg (14 lbs)
Water Tank (ref 91810, empty) 10.43 kg (23 lbs)
Max Scanning Arm Travel 27.5 cm (10.8 in)
Ionization Chamber Holder Diameter Accommodation Max: 20 mm (0.79 in) Min: 6 mm (0.24 in)
Backscatter clearance at 25 cm depth ˜ 8 cm (3.1) including phantom base
Accuracy of Position ± 0.05 mm (0.002 in) over entire 275.00 mm
Repeatability of Position ± 0.05 mm (0.002 in) over entire 275.00 mm
Operating Conditions Pressure: 680 - 800 mm Hg
Temperature: 10 - 40 °C
Relative Humidity: 30 to 75%, non-condensing
Storage Conditions Temperature: -40 to 70 °C
Relative Humidity: 0 to 95%, non-condensing
Cabling: 100 ft extension cable provided (standard RS-232 configuration)
Power Requirements AC Output 12 VDC @ 1.25 A
Optional Detector Brackets PTW Markus® (ref 70850)
PTW Roos® (ref 70852)
Vertical Holder (ref 70851)
Product Standards CE