PET and SPECT Image Performance Evaluation

Phantom for high spatial resolution SPECT and PET systems for performance evaluation (collimator, artifacts, calibration, reconstruction parameters). It can be used to evaluate, for example: center-of-rotation error, non-uniformity artifacts, changes of radius-of-rotation on spatial resolution, reconstruction filters on spatial resolution, attenuation and scatter compensation.

Technical data (can be modified according to individual specifications):

  • Main cylinder:
    • Inside cylinder diameter: 206mm
    • Inside cylinder height: 186mm
    • Cylinder wall thickness: 7mm
  • Cold rods insert
    • Rods diameters: 4.8, 6.4, 7.9, 9.5, 11.1 and 12.7 mm
    • Height of rods: 88mm
  • Cold spheres:
    • Solid sphere diameters: 9.5, 12.7, 15.9, 19.1, 25.4 and 31.8mm
    • Height of center of spheres from base plate: 127mm
  • Optional PET Lid with cylindrical samples:
    • Refillable thin-walled cylinders, diameters: 8, 12, 16 and 25mm
    • Water filled cylinder diameter: 25mm
    • Air filled cylinder diameter: 25mm
    • PTFE solid cylinder diameter: 25mm
    • Cylinders height: 38mm
  • Heavy duty carrying case

 

Product features:

  • Complies with:
    • NEMA Standards Publication (NU 1-2001)Performance Measurements of Scintillation Cameras
    • AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
    • AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
    • ACR–SNM technical standard for diagnostic procedures using radiopharmaceuticals (Res. 5 - 2011)
  • CE certified
  • Manual provides detailed guidelines for carrying out each test, results assessment and registration