Mammography Contrast Detail Phantom
Phantom for testing the performance of mammographic X-ray units. It has 49 holes that generate subtle changes in contrast. It can be used to detect small variations in system performance. Phantom is easy to use - just put it on the surface of the image receptor in the same position as the breast and set-up X-ray tube and the breast compression plate as for a normal study.
Technical data:
- Dimensions: 62.7 x 62.7 x 62.7 mm
- Holes depth (7 columns): 0.853, 0.533, 0.332, 0.208, 0.129, 0.080, 0.50 mm
- Typical contrast for mammographic energies: 6.60, 4.20, 2.60, 1.70, 1.00, 0.65, 0.41
- Diameter of holes (7 rows): 4.292, 2.524, 1.485, 0.873, 0.513, 0.302, 0.177 mm
- Optimized for digital units
- Easy to use, compact and light-weight
- Includes rotating support plate
Product features:
- Complies with:
- IEC 61223-3-2
- “European Guidelines for Quality Assurance in Mammography Screening”
- CE certified
- Manual provides detailed guidelines for carrying out each test, results assessment and registration