Pro-CT Align
Radiology
Diagnomatic
This phantom can be used to check the alignment of the internal and external lasers to the radiographic center of CT and PET/CT units and to verify lateral gantry angle. It can also be used with accelerator units to check vertical and lateral gantry angles, laser alignment and vertical table movement.
Benefits
04-601 - CT laser alignment check version
04-602 - CT laser alignment check and calibration version (with aluminium seeds insert)
04-603 - PET/CT alignment check version (with tungsten ball)
Technical data:
Made of transparent PMMA 150 x 150 x 50 mm
Scribed lines are 1 mm wide and white for ease of viewing
Optional 5.5 mm diameter tungsten ball in the center for sterotactic collimator verification (it is removable for CT applications)
Optional insert containing 5 aluminium seeds 1 mm in diameter and 5 mm long with a chart indicating center to center spacing between seeds
Large 150 x 150 mm sides include:
Centring scribe lines that go across the phantom on one side
Two large holes 9.5 mm in diameter, 106 mm apart centre to centre in the exact centre of a quadrant of the phantom
Small 50 x 150 mm sides include:
Scribed lines through the centre of all four sides- 2 of those sides are also scribed 50 mm from the centre
3 pinholes one at the centre and two 50 mm from the centre
Centre pinholes are 2 mm in diameter and hold a tungsten pin 15 mm long
Centre pinholes are 1.0 mm in diameter
Levelling platform:
Made of transparent PMMA
Dimensions: 150 x 150 x 10 mm
Bubble level and three levelling screws
Product features:
Complies with:
- IEC 61223-3-5
- IEC 61223-2-6CE certified
The manual provides detailed guidelines for carrying out each test, results assessment and registration
Let's Talk
Have a question, need a quote, or want to book a demo? We’re here to help.
info@meditron.ch
+41 22 995 01 91
Place de l'Industrie 2, 1180 Rolle Schönenhofstrasse 19, 8500 Frauenfeld

Pro-CT Align
Radiology
Diagnomatic
This phantom can be used to check the alignment of the internal and external lasers to the radiographic center of CT and PET/CT units and to verify lateral gantry angle. It can also be used with accelerator units to check vertical and lateral gantry angles, laser alignment and vertical table movement.
Benefits
04-601 - CT laser alignment check version
04-602 - CT laser alignment check and calibration version (with aluminium seeds insert)
04-603 - PET/CT alignment check version (with tungsten ball)
Technical data:
Made of transparent PMMA 150 x 150 x 50 mm
Scribed lines are 1 mm wide and white for ease of viewing
Optional 5.5 mm diameter tungsten ball in the center for sterotactic collimator verification (it is removable for CT applications)
Optional insert containing 5 aluminium seeds 1 mm in diameter and 5 mm long with a chart indicating center to center spacing between seeds
Large 150 x 150 mm sides include:
Centring scribe lines that go across the phantom on one side
Two large holes 9.5 mm in diameter, 106 mm apart centre to centre in the exact centre of a quadrant of the phantom
Small 50 x 150 mm sides include:
Scribed lines through the centre of all four sides- 2 of those sides are also scribed 50 mm from the centre
3 pinholes one at the centre and two 50 mm from the centre
Centre pinholes are 2 mm in diameter and hold a tungsten pin 15 mm long
Centre pinholes are 1.0 mm in diameter
Levelling platform:
Made of transparent PMMA
Dimensions: 150 x 150 x 10 mm
Bubble level and three levelling screws
Product features:
Complies with:
- IEC 61223-3-5
- IEC 61223-2-6CE certified
The manual provides detailed guidelines for carrying out each test, results assessment and registration