Pro-CT Dose AAPM TG-200 / Pro-CT Dose ICRU 87
Radiology
Diagnomatic
This phantom is designed according to ICRU 87 report and AAPM TG-200 specifications. The phantom comprises high-density polyethylene (mass density of 0.97 g/cm³) and is 300 mm in diameter. The total length of the phantom is 600 mm. This allows capturing the majority of the scattered radiation and gives more accurate results than standard PMMA cylinders – 150 mm long, 320 mm in diameter. Due to the large mass of the phantom, it is manufactured in three separate sections. In order to keep these sections aligned, alignment pegs are included in the design.
Benefits
Technical data:
Total dimensions: diameter 300 mm, length 600 mm
Made of high-density polyethylene
Built of 3 separate 200 mm sections:
- Cable section (A) contains three holes Ø13.1 mm where detector cable rests during measurements
- Functional section (B) contains three holes Ø13.1 mm for positioning CT detector and Ø44.45 mm, 50.8 mm deep hole for MTF (modular transfer function) calculation
- Blank section (C) is free of any structures that can be used for noise power spectrum (NPS) evaluation2x 12 mm thick polyethylene end cups to keep all modules together
2x 420 mm long polyethylene filler plugs to fill unused holes
Additional polyethylene rod to help center the probe tip in the middle section
12x locating plugs to help in correct phantom assembly
Engraved markings on phantoms for easier positioning
Carrying case with a trolley
Product features:
Complies with:
- ICRU report 87
- AAPM TG-200CE certified
The manual provides detailed guidelines for carrying out each test, results assessment and registration
Find a
Let's Talk
Have a question, need a quote, or want to book a demo? We’re here to help.
info@meditron.ch
+41 22 995 01 91
Place de l'Industrie 2, 1180 Rolle Schönenhofstrasse 19, 8500 Frauenfeld

Pro-CT Dose AAPM TG-200 / Pro-CT Dose ICRU 87
Radiology
Diagnomatic
This phantom is designed according to ICRU 87 report and AAPM TG-200 specifications. The phantom comprises high-density polyethylene (mass density of 0.97 g/cm³) and is 300 mm in diameter. The total length of the phantom is 600 mm. This allows capturing the majority of the scattered radiation and gives more accurate results than standard PMMA cylinders – 150 mm long, 320 mm in diameter. Due to the large mass of the phantom, it is manufactured in three separate sections. In order to keep these sections aligned, alignment pegs are included in the design.
Benefits
Technical data:
Total dimensions: diameter 300 mm, length 600 mm
Made of high-density polyethylene
Built of 3 separate 200 mm sections:
- Cable section (A) contains three holes Ø13.1 mm where detector cable rests during measurements
- Functional section (B) contains three holes Ø13.1 mm for positioning CT detector and Ø44.45 mm, 50.8 mm deep hole for MTF (modular transfer function) calculation
- Blank section (C) is free of any structures that can be used for noise power spectrum (NPS) evaluation2x 12 mm thick polyethylene end cups to keep all modules together
2x 420 mm long polyethylene filler plugs to fill unused holes
Additional polyethylene rod to help center the probe tip in the middle section
12x locating plugs to help in correct phantom assembly
Engraved markings on phantoms for easier positioning
Carrying case with a trolley
Product features:
Complies with:
- ICRU report 87
- AAPM TG-200CE certified
The manual provides detailed guidelines for carrying out each test, results assessment and registration