Pro-MAM ACR FF
Radiology
Diagnomatic
This ACR approved Full Field phantom was designed to test the performance of a digital mammographic system by evaluating the system’s ability to image small structures similar to those found clinically: micro-calcifications, fibrous structures in ducts and tumor-like masses. It is similar to the Pro-MAM ACR phantom, the main difference being the size: the FF (Full Field) version is larger and covers the entire image detector, thus eliminating scatter.
Benefits
Technical data:
Dimensions: 310 x 190 x 41 mm
Simulates 42 mm compressed breast of average glandular / adipose composition (50% / 50%)
Wax insert:
– Nylon fibrils diameters: 0.89, 0.75, 0.61, 0.54, 0.40 and 0.30 mm
– Microcalcifications: 0.32, 0.28, 0.23, 0.20, 0.17 and 0.14 mm specks
– Tumor-like masses: 1.00, 0.75, 0.50, 0.38, 0.25 and 0.20 mm thickComfortable carrying case

Product features:
Complies with:
– IEC 61223-3-2
– ACR Mammography Accreditation Program RequirementsApproved by ACR
CE marked
The manual provides detailed guidelines for carrying out each test, results assessment and registration
Let's Talk
Have a question, need a quote, or want to book a demo? We’re here to help.
info@meditron.ch
+41 22 995 01 91
Place de l'Industrie 2, 1180 Rolle Schönenhofstrasse 19, 8500 Frauenfeld

Pro-MAM ACR FF
Radiology
Diagnomatic
This ACR approved Full Field phantom was designed to test the performance of a digital mammographic system by evaluating the system’s ability to image small structures similar to those found clinically: micro-calcifications, fibrous structures in ducts and tumor-like masses. It is similar to the Pro-MAM ACR phantom, the main difference being the size: the FF (Full Field) version is larger and covers the entire image detector, thus eliminating scatter.
Benefits
Technical data:
Dimensions: 310 x 190 x 41 mm
Simulates 42 mm compressed breast of average glandular / adipose composition (50% / 50%)
Wax insert:
– Nylon fibrils diameters: 0.89, 0.75, 0.61, 0.54, 0.40 and 0.30 mm
– Microcalcifications: 0.32, 0.28, 0.23, 0.20, 0.17 and 0.14 mm specks
– Tumor-like masses: 1.00, 0.75, 0.50, 0.38, 0.25 and 0.20 mm thickComfortable carrying case

Product features:
Complies with:
– IEC 61223-3-2
– ACR Mammography Accreditation Program RequirementsApproved by ACR
CE marked
The manual provides detailed guidelines for carrying out each test, results assessment and registration