Pro-MAM Geometry – Meditron Switzerland

Pro-MAM Geometry

Radiology

Diagnomatic

This test device was developed to provide the user with a convenient, quick and accurate way of testing collimation of mammography units. It is very simple to use. Instead of using all those hard-to-find coins, you only use our test tool.

Speak to an advisor

Franco Quaranta

Radiology

expert

Benefits

Technical data (can be modifi:

  • 240 mm diameter semicircle

  • first half of the test module contains:
    - 6 groups of 10 mm long multi-directional filaments of the following diameters: 0.40, 0.35, 0.30, 0.25, 0.225, 0.20 mm
    - 6 groups of microcalcifications in the following ranges: 354-224, 283-180, 226-150, 177-106, 141-90, 106-63 μm
    - 6 groups of 3 low contrast details, 3mm in diameter, producing nominal contrast at 28 kVp of: 0.04, 0.03, 0.02, 0.015, 0.01, 0.005 (each contrast is presented 3 times)

  • second half of the test module contains:
    - clinically realistic breast tissue (not tissue equivalent to but simulating the appearance of the breast tissue) containing 6 groups of low contrast details with the same specifications as in the other half

  • 6 PMMA plates 10 mm thick

  • 2 PMMA plates 5 mm thick

  • carrying case

‍

Product features:

  • complies with:
    – IEC 61223-3-2

  • this is not an ACR approved phantom for accreditation

  • CE marked

  • the manual provides detailed guidelines for carrying out each test, results assessment and registration

Let's Talk

Have a question, need a quote, or want to book a demo? We’re here to help.

info@meditron.ch

+41 22 995 01 91

Place de l'Industrie 2, 1180 Rolle

Schönenhofstrasse 19, 8500 Frauenfeld

Pro-MAM Geometry – Meditron Switzerland

Pro-MAM Geometry

Radiology

Diagnomatic

This test device was developed to provide the user with a convenient, quick and accurate way of testing collimation of mammography units. It is very simple to use. Instead of using all those hard-to-find coins, you only use our test tool.

Speak to an advisor

Franco Quaranta

Radiology

expert

Benefits

Technical data (can be modifi:

  • 240 mm diameter semicircle

  • first half of the test module contains:
    - 6 groups of 10 mm long multi-directional filaments of the following diameters: 0.40, 0.35, 0.30, 0.25, 0.225, 0.20 mm
    - 6 groups of microcalcifications in the following ranges: 354-224, 283-180, 226-150, 177-106, 141-90, 106-63 μm
    - 6 groups of 3 low contrast details, 3mm in diameter, producing nominal contrast at 28 kVp of: 0.04, 0.03, 0.02, 0.015, 0.01, 0.005 (each contrast is presented 3 times)

  • second half of the test module contains:
    - clinically realistic breast tissue (not tissue equivalent to but simulating the appearance of the breast tissue) containing 6 groups of low contrast details with the same specifications as in the other half

  • 6 PMMA plates 10 mm thick

  • 2 PMMA plates 5 mm thick

  • carrying case

‍

Product features:

  • complies with:
    – IEC 61223-3-2

  • this is not an ACR approved phantom for accreditation

  • CE marked

  • the manual provides detailed guidelines for carrying out each test, results assessment and registration