Benefits
Technical data:
240 mm diameter semicircle
First half of the test module contains:
- 6 groups of 10 mm long multi-directional filaments of the following diameters: 0.40, 0.35, 0.30, 0.25, 0.225, 0.20 mm
- 6 groups of microcalcifications in the following ranges: 354-224, 283-180, 226-150, 177-106, 141-90, 106-63 μm
- 6 groups of 3 low contrast details, 3mm in diameter, producing nominal contrast at 28 kVp of: 0.04, 0.03, 0.02, 0.015, 0.01, 0.005 (each contrast is presented 3 times)Second half of the test module contains:
- clinically realistic breast tissue (not tissue equivalent to but simulating the appearance of the breast tissue) containing 6 groups of low contrast details with the same specifications as in the other half6 PMMA plates 10 mm thick
2 PMMA plates 5 mm thick
Carrying case
Product features:
Complies with:
– IEC 61223-3-2This is not an ACR approved phantom for accreditation
CE marked
The manual provides detailed guidelines for carrying out each test, results assessment and registration
Let's Talk
Have a question, need a quote, or want to book a demo? We’re here to help.
info@meditron.ch
+41 22 995 01 91
Place de l'Industrie 2, 1180 Rolle Schönenhofstrasse 19, 8500 Frauenfeld
Benefits
Technical data:
240 mm diameter semicircle
First half of the test module contains:
- 6 groups of 10 mm long multi-directional filaments of the following diameters: 0.40, 0.35, 0.30, 0.25, 0.225, 0.20 mm
- 6 groups of microcalcifications in the following ranges: 354-224, 283-180, 226-150, 177-106, 141-90, 106-63 μm
- 6 groups of 3 low contrast details, 3mm in diameter, producing nominal contrast at 28 kVp of: 0.04, 0.03, 0.02, 0.015, 0.01, 0.005 (each contrast is presented 3 times)Second half of the test module contains:
- clinically realistic breast tissue (not tissue equivalent to but simulating the appearance of the breast tissue) containing 6 groups of low contrast details with the same specifications as in the other half6 PMMA plates 10 mm thick
2 PMMA plates 5 mm thick
Carrying case
Product features:
Complies with:
– IEC 61223-3-2This is not an ACR approved phantom for accreditation
CE marked
The manual provides detailed guidelines for carrying out each test, results assessment and registration