Pro-NM NEMA 94
Radiology
Diagnomatic
Main application of Pro-NM NEMA 94 is PET acceptance testing with NEMA standard. Evaluation of count rate, uniformity, scatter fraction, attenuation compensation, and scatter compensation of ECT systems. The Pro-NM NEMA 94 is designed in accordance with the recommendations by the National Electrical Manufacturers Association (NEMA) to standardize the measurement of performance of PET.
Benefits
Technical data:
Cylinder outside diameter: 200 mm
Cylinder insider diameter: 194 mm
Wall thickness: 3 mm
Teflon Insert diameter: 51 mm
Fillable Insert inside diameter: 45 mm
Fillable Insert height: 185 mm
Fillable Insert volume: ~ 295 mL
Line Source diameter: ~ 1 mm
Line Source height: 184 mm
Product features:
Complies with:
- Performance Measurements of Positron Emission Tomographs, NEMA Standards Publication NU2, National Electrical Manufacturers Association (NEMA), Washington, D.C., 1994.
- Karp JS, Daube-Witherspoon ME, Hoffman EJ et al. “Performance standards in positron emission tomography”, J Nucl Med, 32:2342-2350.CE certified
The manual provides detailed guidelines for carrying out each test, results assessment and registration
Let's Talk
Have a question, need a quote, or want to book a demo? We’re here to help.
info@meditron.ch
+41 22 995 01 91
Place de l'Industrie 2, 1180 Rolle Schönenhofstrasse 19, 8500 Frauenfeld

Pro-NM NEMA 94
Radiology
Diagnomatic
Main application of Pro-NM NEMA 94 is PET acceptance testing with NEMA standard. Evaluation of count rate, uniformity, scatter fraction, attenuation compensation, and scatter compensation of ECT systems. The Pro-NM NEMA 94 is designed in accordance with the recommendations by the National Electrical Manufacturers Association (NEMA) to standardize the measurement of performance of PET.
Benefits
Technical data:
Cylinder outside diameter: 200 mm
Cylinder insider diameter: 194 mm
Wall thickness: 3 mm
Teflon Insert diameter: 51 mm
Fillable Insert inside diameter: 45 mm
Fillable Insert height: 185 mm
Fillable Insert volume: ~ 295 mL
Line Source diameter: ~ 1 mm
Line Source height: 184 mm
Product features:
Complies with:
- Performance Measurements of Positron Emission Tomographs, NEMA Standards Publication NU2, National Electrical Manufacturers Association (NEMA), Washington, D.C., 1994.
- Karp JS, Daube-Witherspoon ME, Hoffman EJ et al. “Performance standards in positron emission tomography”, J Nucl Med, 32:2342-2350.CE certified
The manual provides detailed guidelines for carrying out each test, results assessment and registration