Pro-NM NEMA NU 2 Resolution
Radiology
Diagnomatic
A phantom for evaluation of spatial resolution of positron emission tomographs (PET). It is used to characterize the widths of the reconstructed image point spread functions (PSF) of compact radioactive sources. It has been designed in accordance with the NEMA Nu 2-2012.
Benefits
Technical data:
Five capillaries with an inside diameter of 1 mm and an outside diameter of less than 2 mm
In the transverse direction the capillary is positioned at 1 cm (to represent the center of the FOV, but positioned to avoid any possible inconsistent results at the very center of the FOV), 10 cm, and 20 cm from the center of the plane
Middle of the capillaries is marked to provide guidance while the radionuclide is introduced
Markers and convenient base with a spirit level for easy and accurate positioning
Optional carrying case (08-109-CS)
Product features:
Complies with:
- International Standard: Radionuclide imaging devices – Characteristics and test conditions – Part 1: Positron emission tomographs, International Electrotechnical Commission (IEC), 61675-1, Geneva, Switzerland, 1998.
- Performance Measurements of Scintillation Cameras, NEMA Standards Publication No. NU2, National Electrical Manufacturers Association (NEMA), Washington, D.C., 2001.
- NEMA 2007/IEC2008
- NEMA 2012/IEC 2008
- NEMA Standard NU 2-2018CE certified
The manual provides detailed guidelines for carrying out each test, results assessment and registration
Let's Talk
Have a question, need a quote, or want to book a demo? We’re here to help.
info@meditron.ch
+41 22 995 01 91
Place de l'Industrie 2, 1180 Rolle Schönenhofstrasse 19, 8500 Frauenfeld

Pro-NM NEMA NU 2 Resolution
Radiology
Diagnomatic
A phantom for evaluation of spatial resolution of positron emission tomographs (PET). It is used to characterize the widths of the reconstructed image point spread functions (PSF) of compact radioactive sources. It has been designed in accordance with the NEMA Nu 2-2012.
Benefits
Technical data:
Five capillaries with an inside diameter of 1 mm and an outside diameter of less than 2 mm
In the transverse direction the capillary is positioned at 1 cm (to represent the center of the FOV, but positioned to avoid any possible inconsistent results at the very center of the FOV), 10 cm, and 20 cm from the center of the plane
Middle of the capillaries is marked to provide guidance while the radionuclide is introduced
Markers and convenient base with a spirit level for easy and accurate positioning
Optional carrying case (08-109-CS)
Product features:
Complies with:
- International Standard: Radionuclide imaging devices – Characteristics and test conditions – Part 1: Positron emission tomographs, International Electrotechnical Commission (IEC), 61675-1, Geneva, Switzerland, 1998.
- Performance Measurements of Scintillation Cameras, NEMA Standards Publication No. NU2, National Electrical Manufacturers Association (NEMA), Washington, D.C., 2001.
- NEMA 2007/IEC2008
- NEMA 2012/IEC 2008
- NEMA Standard NU 2-2018CE certified
The manual provides detailed guidelines for carrying out each test, results assessment and registration