Pro-NM Performance – Meditron Switzerland

Pro-NM Performance

Radiology

Diagnomatic

The phantom for NM and PET systems performance evaluation (collimator, artifacts, calibration, reconstruction parameters). It can be used to evaluate, for example: center-of-rotation error, non-uniformity artifacts, changes of radius-of-rotation on spatial resolution, reconstruction filters on spatial resolution, attenuation and scatter compensation.

Speak to an advisor

Franco Quaranta

Radiology

expert

Benefits

08-101 - standard version
08-103 - version with the PET Lid

Technical data:

Main cylinder

  • Inside cylinder diameter: 206 mm

  • Inside cylinder height: 186 mm

  • Cylinder wall thickness: 7 mm

Cold rods insert

  • Rod diameters: 4.8, 6.4, 7.9, 9.5, 11.1 and 12.7 mm

  • Height of rods: 88 mm

Cold spheres

  • Solid sphere diameters: 9.5, 12.7, 15.9, 19.1, 25.4 and 31.8 mm

  • Height of the center of the spheres from the base plate: 127 mm

Optional PET Lid with cylindrical samples (08-102)

  • Refillable thin-walled cylinders, diameters: 8, 12, 16 and 25 mm

  • Water filled cylinder diameter: 25 mm

  • Air filled cylinder diameter: 25 mm

  • PTFE solid cylinder diameter: 25 mm

  • Cylinder height: 38 mm

Accessories

  • Optional heavy duty carrying case (08-101-CS)


Product features:

  • Complies with:
    - NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
    - AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
    - AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
    - ACR–SNM  (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals

  • CE certified

  • The manual provides detailed guidelines for carrying out each test, results assessment and registration

Let's Talk

Have a question, need a quote, or want to book a demo? We’re here to help.

info@meditron.ch

+41 22 995 01 91

Place de l'Industrie 2, 1180 Rolle

Schönenhofstrasse 19, 8500 Frauenfeld

Pro-NM Performance – Meditron Switzerland

Pro-NM Performance

Radiology

Diagnomatic

The phantom for NM and PET systems performance evaluation (collimator, artifacts, calibration, reconstruction parameters). It can be used to evaluate, for example: center-of-rotation error, non-uniformity artifacts, changes of radius-of-rotation on spatial resolution, reconstruction filters on spatial resolution, attenuation and scatter compensation.

Speak to an advisor

Franco Quaranta

Radiology

expert

Benefits

08-101 - standard version
08-103 - version with the PET Lid

Technical data:

Main cylinder

  • Inside cylinder diameter: 206 mm

  • Inside cylinder height: 186 mm

  • Cylinder wall thickness: 7 mm

Cold rods insert

  • Rod diameters: 4.8, 6.4, 7.9, 9.5, 11.1 and 12.7 mm

  • Height of rods: 88 mm

Cold spheres

  • Solid sphere diameters: 9.5, 12.7, 15.9, 19.1, 25.4 and 31.8 mm

  • Height of the center of the spheres from the base plate: 127 mm

Optional PET Lid with cylindrical samples (08-102)

  • Refillable thin-walled cylinders, diameters: 8, 12, 16 and 25 mm

  • Water filled cylinder diameter: 25 mm

  • Air filled cylinder diameter: 25 mm

  • PTFE solid cylinder diameter: 25 mm

  • Cylinder height: 38 mm

Accessories

  • Optional heavy duty carrying case (08-101-CS)


Product features:

  • Complies with:
    - NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
    - AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
    - AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
    - ACR–SNM  (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals

  • CE certified

  • The manual provides detailed guidelines for carrying out each test, results assessment and registration