Pro-NM Performance MINI
Radiology
Diagnomatic
The phantom for performance evaluation of small animal systems or ultra-high resolution ECT systems. It can be used to evaluate: center-of-rotation error, non-uniformity artifacts and spatial resolution. Available in two sizes for FOV at least 45 mm or 75 mm.
Benefits
At least 75 mm FOV version:
Main cylinder:
Outside cylinder diameter: 80 mm
Inside cylinder diameter: 74 mm
Inside cylinder height: 65 mm
Cold rods insert:
Rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
Height of rods: 34 mm
At least 45 mm FOV version:
Main cylinder:
Outside cylinder diameter: 50 mm
Inside cylinder diameter: 45 mm
Inside cylinder height: 65 mm
Cold rods insert:
Rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
Height of rods: 34 mm
Accessories
Optional carrying case
Product features:
Complies with:
- NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
- AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
- AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
- ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticalsCE certified
The manual provides detailed guidelines for carrying out each test, results assessment and registration
Let's Talk
Have a question, need a quote, or want to book a demo? We’re here to help.
info@meditron.ch
+41 22 995 01 91
Place de l'Industrie 2, 1180 Rolle Schönenhofstrasse 19, 8500 Frauenfeld

Pro-NM Performance MINI
Radiology
Diagnomatic
The phantom for performance evaluation of small animal systems or ultra-high resolution ECT systems. It can be used to evaluate: center-of-rotation error, non-uniformity artifacts and spatial resolution. Available in two sizes for FOV at least 45 mm or 75 mm.
Benefits
At least 75 mm FOV version:
Main cylinder:
Outside cylinder diameter: 80 mm
Inside cylinder diameter: 74 mm
Inside cylinder height: 65 mm
Cold rods insert:
Rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
Height of rods: 34 mm
At least 45 mm FOV version:
Main cylinder:
Outside cylinder diameter: 50 mm
Inside cylinder diameter: 45 mm
Inside cylinder height: 65 mm
Cold rods insert:
Rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
Height of rods: 34 mm
Accessories
Optional carrying case
Product features:
Complies with:
- NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
- AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
- AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
- ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticalsCE certified
The manual provides detailed guidelines for carrying out each test, results assessment and registration