Pro-RF Fluo 18
Radiology
Diagnomatic
The phantom for evaluation of imaging performance of fluoroscopic systems. It allows brightness and contrast adjustments, circulargeometry check (scanning linearity) plus wide-range low-contrast and high-contrast resolution evaluation. Available in a standard layouts plus any custom configuration.
Benefits
Technical data:
Diameter: 180 mm
Thickness: 10 mm
Two standard layouts available (others upon request)
3 circular rims on the outside of the phantom (Ø: 150, 160 and 170 mm) for a geometry and collimation check
18 low contrast objects, 8 mm in diameter, producing contrast values from 0.9 to 16.7%
Pattern for line pair resolution evaluation (from 0.6 to 5.0 LP/mm)
Pb and Cu squares with circular bright and dark objects for contrast and brightness adjustment
1.0 and 0.5 mm high purity Cu attenuation filters (fit to the diaphragm housing)
Compact carrying case
Product features:
Complies with:
– IEC 61223-3-1– IPEM Report 91 (2005): Recommended Standards for the Routine Performance Testing of Diagnostic X-Ray Imaging Systems published by the Institute of Physics and Engineering in MedicineCE marked
The manual provides detailed guidelines for carrying out each test, results assessment and registration
Let's Talk
Have a question, need a quote, or want to book a demo? We’re here to help.
info@meditron.ch
+41 22 995 01 91
Place de l'Industrie 2, 1180 Rolle Schönenhofstrasse 19, 8500 Frauenfeld

Pro-RF Fluo 18
Radiology
Diagnomatic
The phantom for evaluation of imaging performance of fluoroscopic systems. It allows brightness and contrast adjustments, circulargeometry check (scanning linearity) plus wide-range low-contrast and high-contrast resolution evaluation. Available in a standard layouts plus any custom configuration.
Benefits
Technical data:
Diameter: 180 mm
Thickness: 10 mm
Two standard layouts available (others upon request)
3 circular rims on the outside of the phantom (Ø: 150, 160 and 170 mm) for a geometry and collimation check
18 low contrast objects, 8 mm in diameter, producing contrast values from 0.9 to 16.7%
Pattern for line pair resolution evaluation (from 0.6 to 5.0 LP/mm)
Pb and Cu squares with circular bright and dark objects for contrast and brightness adjustment
1.0 and 0.5 mm high purity Cu attenuation filters (fit to the diaphragm housing)
Compact carrying case
Product features:
Complies with:
– IEC 61223-3-1– IPEM Report 91 (2005): Recommended Standards for the Routine Performance Testing of Diagnostic X-Ray Imaging Systems published by the Institute of Physics and Engineering in MedicineCE marked
The manual provides detailed guidelines for carrying out each test, results assessment and registration